Series KBWF | Growth chambers with light and humidity

code : KBWF240

Important characteristics

  • Temperature range: 0 °C to 70 °C
  • Temperature range with light: 10 °C to 60 °C
  • Humidity range: 10 % to 80 % RH
  • 2 positionable illumination cassettes, each with 5 fluorescent tubes
  • APT.line™ preheating chamber technology
  • Adjustable fan speed
  • Humidity regulation with capacitative humidity sensor and vapor humidification
  • Intuitive touchscreen controller with time-segment and real-time programming
  • Internal data logger, measured values can be read out in open format via USB
  • Unit self-test for comprehensive status analysis
  • Tight-sealing inner door made of safety glass (ESG)
  • Avoidance of glass corrosion by special TIMELESS coating
  • Display via color LCD monitor
  • 2 stainless steel racks
  • Access port with silicone plug, 30 mm, left
  • Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
  • Computer interface: Ethernet
  • Door heating

Select version
Rated voltage Option model Version Art. No. MSRP*[?] Selection Bookmark
Model KBWF 240
200…240 V 1~ 50 Hz

Standard

KBWF240-230V¹ 9020-0336 - +
* Recommended retail price

 

Designation KBWF240-230V¹
Option model Standard
Order number 9020-0336
Measures
Interior volume [L] 247
Net weight of the unit (empty) [kg] 214
permitted load [kg] 100
Load per rack [kg] 30
Doors
Inner doors 1
Unit doors 1
Internal Dimensions
Width [mm] 650
Height [mm] 785
Depth [mm] 485
Outer dimensions
Width net [mm] 925
Height net [mm] 1465
Depth net [mm] 800
Wall clearance sidewise [mm] 100
Wall clearance back [mm] 100
Environment-specific data
Energy consumption at 37 °C and 75 % RH [Wh/h] 760
Sound-pressure level [dB(A)] 53
Fixtures
Number of shelves (std./max.) 2/7
Number of illumination cassettes (std./max.) 2/2
Performance Data Temperature
Temperature range without illumination cassettes [°C] 0…70
Temperature range with 100% illumination [°C] 10…60
Performance Data Climate
Temperature range without illumination cassettes [°C] 10…70
Temperature range with 100% illumination [°C] 10…60
Humidity range without illumination [% RH] 10…80
Humidity range with illumination cassettes, with illumination [% RH] 10…75
Lightdata per Illumination Cassettes
Daylight tubes [W/m²]* 24
Daylight tubes [lx]* 9000
Fluora® growth lamps [W/m²]* 23
Fluora® growth lamps [lx]* 7500
Arabidopsis lamps [W/m²]* 32
Arabidopsis lamps [lx]* 11000
Electrical data
Rated Voltage [V] 200…240
Power frequency [Hz] 50
Nominal power [kW] 0
Unit fuse [A] 16
Phase (Nominal voltage) 1~
  1. All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.
* Average value, measured at 25 °C with a spherical sensor (±10 %) by 12 cm below the light cassette. The values given in W/m2 refer to global radiation.

Accessories Series KBWF | Growth chambers with light and humidity