Series ED 23 | Drying and heating chambers Classic.Line with gravity convection

code : ED23

Important characteristics

  • Temperature range: room temperature plus 5 °C to 300 °C
  • APT.line™ preheating chamber technology
  • Gravity convection
  • Adjustable exhaust air flap
  • Controller with timer functions
  • 2 chrome-plated racks
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm

Select version
Rated voltage Option model Version Art. No. MSRP*[?] Selection Bookmark
Model ED 23
120 V 1~ 60 Hz

Standard

ED023UL-120V¹ 9010-0192 - +

with RS 422

ED023UL-120V-I¹ 9010-0193 - +
230 V 1~ 50/60 Hz

Standard

ED023-230V¹ 9010-0190 - +

with RS 422

ED023-230V-I¹ 9010-0191 - +
* Recommended retail price

 

Designation ED023UL-120V¹ ED023UL-120V-I¹ ED023-230V¹ ED023-230V-I¹
Option model Standard with RS 422 Standard with RS 422
Order number 9010-0192 9010-0193 9010-0190 9010-0191
Measures
Interior volume [L] 20 20 20 20
Net weight of the unit (empty) [kg] 27 27 27 27
permitted load [kg] 25 25 25 25
Load per rack [kg] 12 12 12 12
Doors
Unit doors 1 1 1 1
Internal Dimensions
Width [mm] 222 222 222 222
Height [mm] 330 330 330 330
Depth [mm] 300 300 300 300
Outer dimensions
Width net [mm] 435 435 435 435
Height net [mm] 495 495 495 495
Depth net [mm] 520 520 520 520
Wall clearance back [mm] 100 100 100 100
Wall clearance sidewise [mm] 100 100 100 100
Environment-specific data
Energy consumption at 150 °C [Wh/h] 148 148 148 148
Fixtures
Number of shelves (std./max.) 2/4 2/4 2/4 2/4
Data
  - - - -
Online
Shop URL https://www.binder-shop.com/us/oven/oven-with-gravity-convection-ed023ul-120v - - -
Performance Data Temperature
Temperature range 5 °C above ambient temperature to [°C] 300 300 300 300
Temperature variation at 150 °C [± K] 2.8 2.8 2.8 2.8
Temperature fluctuation at 150 °C [± K] 0.5 0.5 0.5 0.5
Heating-up time to 150 °C [min] 28 28 28 28
Recovery time after 30 seconds door open at 150 °C [min] 28 28 28 28
Air change data
Air change (approx.) at 150 °C [x/h] 13 13 13 13
Electrical data
Rated Voltage [V] 120 120 230 230
Power frequency [Hz] 60 60 50/60 50/60
Nominal power [kW] 0.8 0.8 0.8 0.8
Unit fuse [A] 12.5 12.5 10 10
Phase (Nominal voltage) 1~ 1~ 1~ 1~
  1. All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10%. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.

Accessories Series ED 23 | Drying and heating chambers Classic.Line with gravity convection