• +2-35734687

  • sales@technoscientific.net

Stay Connected:

Series FDL 115 | Safety drying chambers for limited quantities of solvent

code : FDL115

Important characteristics

  • Temperature range: room temperature plus 5 °C to 300 °C
  • APT.line™ preheating chamber technology
  • Controller with time-segment and real-time programming
  • 2 chrome-plated racks
  • Fresh-air monitoring with audible and visual alarm and automatic deactivation of heating
  • Replaceable filter cartridge, Class M6
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • Computer interface: RS 422

Select version
Rated voltage Option model Version Art. No. MSRP*[?] Selection Bookmark
Model FDL 115
230 V 1~ 50/60 Hz

Standard

FDL115-230V¹ 9010-0292 - +
* Recommended retail price
Designation FDL115-230V¹
Option model Standard
Order number 9010-0292
Measures
Interior volume [L] 115
Net weight of the unit (empty) [kg] 90
permitted load [kg] 60
Load per rack [kg] 30
Internal Dimensions
Width [mm] 600
Height [mm] 435
Depth [mm] 435
Outer dimensions
Width net [mm] 830
Height net [mm] 805
Depth net [mm] 685
Wall clearance back [mm] 160
Wall clearance sidewise [mm] 100
Environment-specific data
Energy consumption at 150 °C [Wh/h] 1200
Sound-pressure level [dB(A)] 57
Fixtures
Number of shelves (std./max.) 2/5
Performance Data Temperature
Temperature range 10 °C above ambient temperature to [°C] 300
Temperature variation at 150 °C [± K] 2.5
Temperature fluctuation [± K] 0,8 ± K
Heating-up time to 150 °C [min] 20
Recovery time after 30 seconds door open at 150 °C [min] 12
Air change data
Air circulation (approx.) [x/min] 20
Volumetric flow rate of exhaust air acc. to EN 1539 at 50 °C [ca. L/min] 400
solvent quantity
Highest permitted solvent quantity (at T-180 °C, M-100g/mol, U-40g/m3, K=0,5) [g] 6.65
Electrical data
Rated Voltage [V] 230
Power frequency [Hz] 50/60
Nominal power [kW] 2.9
Phase (Nominal voltage) 1~
  1. All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.

Accessories Series FDL 115 | Safety drying chambers for limited quantities of solvent