• +2-35734687

  • sales@technoscientific.net

Stay Connected:

Series VDL 23 | Vacuum drying chambers for flammable solvents

code : VDL23

Important characteristics

  • Temperature range: room temperature plus 15 °C to 200 °C
  • ATEX compliance chamber's interior: Ex II -/3G c IIB T3-T1 Gc X
  • APT.line™ preheating chamber technology
  • Controller with time-segment and real-time programming
  • 2 aluminum expansion racks
  • Inert gas connection
  • Shatterproof, spring-mounted safety glass panel
  • Pressurized instrument panel with compressed air connection
  • Pressure control device for heating activated < 125 mbar
  • Flame protection gasket
  • Class 2 independent adjustable temperature safety device (DIN 12880) with visual alarm
  • Computer interface: RS 422

Select version
Rated voltage Option model Version Art. No. MSRP*[?] Selection Bookmark
Model VDL 23
230 V 1~ 50/60 Hz

Standard

VDL023-230V¹ 9030-0038 - +
* Recommended retail price
Designation VDL023-230V¹
Option model Standard
Order number 9030-0038
Measures
Interior volume [L] 23
Net weight of the unit (empty) [kg] 63
permitted load [kg] 35
Load per rack [kg] 20
Doors
Unit doors 1
Internal Dimensions
Width [mm] 285
Height [mm] 285
Depth [mm] 295
Outer dimensions
Width net [mm] 515
Height net [mm] 655
Depth net [mm] 500
Wall clearance back [mm] 100
Wall clearance sidewise [mm] 135
Viewing window width [mm] 250
Viewing window height [mm] 207
Fixtures
Number of shelves (std./max.) 2/4
Performance Data Temperature
Temperature range 15 °C above ambient temperature to [°C] 200
Temperature variation at 100 °C [± K] 1.5
Temperature fluctuation [± K] 0,1 ± K
Heating-up time to 100 °C [min] 65
Vacuum Data
Permitted end vacuum [mbar] 0.01
Leak rate [bar/h] 0.01
Electrical data
Rated Voltage [V] 230
Power frequency [Hz] 50/60
Nominal power [kW] 0.8
Unit fuse [A] 10
Phase (Nominal voltage) 1~
Connections
Vacuum connection with small flange [DN mm] 16
Measuring access port with small flange [DN mm] 16
Inert gas connection with flow limiter (RP") 3/8
Compressed air connection for pressure-encapsulation [mm] 8
  1. All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10%. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.

Accessories Series VDL 23 | Vacuum drying chambers for flammable solvents