• +2-35734687

  • sales@technoscientific.net

Stay Connected:

BD 600 GLP

code : BD 600 GLP

Highlights

  • Good Laboratory Practice compliant
  • For OECD 301F and BOD testing
  • Protected life time data memory (1 GB)
  • Extended measurement duration up to 90 days possible

The BD 600 is a respirometric system for the determination of Biochemical Oxygen Demand (BOD). Save time, reduce the potential for errors and easily interpret data to help make process control decisions for your plant.

Applications

  • Waste Water Treatment

The BD 600 is the ideal system for process control or as a secondary test to the Dilution Method. Using a respirometric system solves many of the issues associated with the dilution method for BOD analysis.

  • Significantly Reduced Sample Preparation 
    It's easy to begin collecting data - simply add sample to the bottle, add nitrification inhibitor and place the sensor head on the bottle. There is no need for sample dilutions or seeding the sample. Because this is a respirometric test, concerns regarding air in the sample are eliminated.
  • Operators do not have to be present at completion of test 
    Measurements are automatically taken and stored at regular time intervals and can be recalled at your convenience - no more weekend trips to the lab or plant!  
  • Easy and quick interpretation of measurement values 
    In addition to displaying results of a sample directly in mg/l BOD, the large, integrated, display is able to display the BOD curve which makes it is easy to see and understand measurement data points and trends. In addition, each sensor head has an integrated LED. This allows operators to visually confirm which sample the data is being displayed for and reduces potential confusion or errors over which sample data is presented
  • Consistent Measurement Conditions 
    The Automatic Start Function ensures all measurements are taken at an equal sample temperature. When the measurement conditions are consistent across readings, operators do not have to work about temperature or pressure fluctuations skewing results.
  • Sample Identification and Tracking 
    The BD 600 interface, makes it easy to assign an identification code or sample name when setting up a new sample. This helps track a sample from the time it is collected and ensures traceability during sample measurement.  
  • Use the remote control 
    If using an incubator with a glass door, operators can use the supplied remote control to display measurement data for a sample - without even opening the door! This further ensures that samples stay at a consistent temperature which is essential to accurate results.
Display backlit
Interfaces USB
External Storage USB
Internal Storage 1 GB or up to 50000 measurement or calibration series
Power Supply 100 - 240 V / 50-60 Hz
3 alkali-manganese batteries (Baby cells/ size C)
Auto – OFF No
Measuring Principle Manometric; mercury-free; electronic pressure sensor
Measuring Range 0-40, 0-80, 0-200, 0-400, 0-800, 0-2000, 0-4000 mg/l
Clock Real Time Clock and Date
Measurement Time User-selectable, between 1 and 28 days
Portability Benchtop
Autostart false
Compliance CE
EMC according to DIN EN 61326
Protection Class IP 53
Dimensions 181 x 230 x 375 mm
Weight (4100 g, unit with bottles & batteries 5775 g, complete with stirring unit)
Weight with Packaging (4100 g, unit with bottles & batteries 5775 g, complete with stirring unit)

Accessories BD 600 GLP