Series KB | Refrigerated incubators with compressor technology

code : KB240

Important characteristics

  • Temperature range (Model 23): 0 °C to 100 °C
  • Temperature range (Models 53, 115, 240, 400, 720): -5 °C to 100 °C
  • APT.line™ preheating chamber technology
  • Refrigeration with compressor refrigerated unit
  • Door heating system from 240 liters
  • Adjustable fan speed
  • Controller with time-segment and real-time programming
  • Inner door made of tempered safety glass
  • 2 stainless steel racks
  • Units up to 115 liters are stackable
  • Access port with silicone plug from 240 liters
  • Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
  • Computer interface: Ethernet

Select version
Rated voltage Option model Version Art. No. MSRP*[?] Selection Bookmark
Model KB 240
200…240 V 1~ 50 Hz

Standard

KB240-230V¹ 9020-0202 - +
100…120 V 1~ 60 Hz

Standard

KB240UL-120V¹ 9020-0304 - +
* Recommended retail price

 

Designation KB240UL-120V¹ KB240-230V¹
Option model Standard Standard
Order number 9020-0304 9020-0202
Measures
Interior volume [L] 247 247
Net weight of the unit (empty) [kg] 171 171
permitted load [kg] 100 100
Load per rack [kg] 30 30
Doors
Inner doors 1 1
Unit doors 1 1
Internal Dimensions
Width [mm] 650 650
Height [mm] 785 785
Depth [mm] 485 485
Outer dimensions
Width net [mm] 925 925
Height net [mm] 1465 1465
Depth net [mm] 800 800
Wall clearance sidewise [mm] 100 100
Wall clearance back [mm] 100 100
Environment-specific data
Energy consumption at 37 °C [Wh/h] 270 270
Sound-pressure level [dB(A)] 53 53
Fixtures
Number of shelves (std./max.) 2/9 2/9
Online
Shop URL https://www.binder-shop.com/us/incubators/refrigerated-incubators/refrigerated-incubator-with-compressor-technology-kb240ul-120v -
Performance Data Temperature
Temperature range [°C] -5…100 -5…100
Temperature variation at 37 °C [± K] 0.2 0.2
Temperature fluctuation at 37 °C [± K] 0,1 ± K 0,1 ± K
Recovery time after 30 seconds door open at 37 °C [min] 2 2
Electrical data
Rated Voltage [V] 100…120 200…240
Power frequency [Hz] 60 50
Nominal power [kW] 1.2 0
Unit fuse [A] 16 16
Phase (Nominal voltage) 1~ 1~
  1. All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.