Series KBF LQC 240 | Humidity test chambers with ICH-compliant light source and light dose control

code : KBF LQC240

Important characteristics

  • Temperature range: 0 °C to 70 °C
  • Temperature range with light: 10 °C to 60 °C
  • Humidity range: 10 % to 80 % RH
  • 2 positionable illumination cassettes with ICH-compliant UV/Vis light source
  • Independent light-dose control of UV-A and visible light with spherical sensors
  • APT.line™ preheating chamber technology
  • Humidity regulation with capacitive humidity sensor and vapor humidification
  • Inner chamber made of stainless steel
  • Intuitive touchscreen controller with time-segment and real-time programming
  • Internal data logger, measured values can be read out in open format via USB
  • Unit self-test for comprehensive status analysis
  • Tight-sealing inner door made of safety glass (ESG)
  • Avoidance of glass corrosion by special TIMELESS coating
  • 2 stainless steel racks
  • Access port with silicone plug, 30 mm, left
  • 4 stable castors, two with brakes
  • Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
  • Computer interface: Ethernet
  • Door heating

Select version
Rated voltage Option model Version Art. No. MSRP*[?] Selection Bookmark
Model KBF LQC 240
200…240 V 1~ 50/60 Hz

Standard

KBFLQC240UL-240V¹ 9020-0333 - +
200…230 V 1~ 50/60 Hz

Standard

KBFLQC240-230V¹ 9020-0332 - +
* Recommended retail price
Designation KBFLQC240UL-240V¹ KBFLQC240-230V¹
Option model Standard Standard
Order number 9020-0333 9020-0332
Measures
Interior volume [L] 247 247
Net weight of the unit (empty) [kg] 213 213
permitted load [kg] 100 100
Load per rack [kg] 30 30
Doors
Inner doors 1 1
Unit doors 1 1
Internal Dimensions
Width [mm] 650 650
Height [mm] 785 785
Depth [mm] 485 485
Outer dimensions
Width net [mm] 925 925
Height net [mm] 1460 1460
Depth net [mm] 800 800
Wall clearance sidewise [mm] 100 100
Wall clearance back [mm] 100 100
Environment-specific data
Energy consumption at 40 °C and 75 % RH [Wh/h] 760 760
Sound-pressure level [dB(A)] 53 53
Fixtures
Number of shelves (std./max.) 2/7 2/7
 
Net weight of the unit (empty) [lbs] 01 01
Fixtures
Number of illumination cassettes (std./max.) 2/2 2/2
Performance Data Temperature
Temperature range without illumination cassettes [°C] 0…70 0…70
Temperature range with 100% illumination [°C] 10…60 10…60
Max. heat compensation at 40 °C with illumination [W] 400 400
Performance Data Climate
Temperature range with 100% illumination [°C] 10…60 10…60
Temperature range without illumination cassettes [°C] 10…70 10…70
Temperature variation with illumination at 40 °C and 75 % RH [± K] 0.6 0.6
Temperature variation with illumination at 25 °C and 60 % RH [± K] 0.6 0.6
Temperature fluctuation with illumination at 25 °C and 60 % RH [± K] 0.2 0.2
Temperature fluctuation with illumination at 40 °C and 75 % RH [± K] 0.2 0.2
Humidity range with illumination cassettes, with illumination [% RH] 10…75 10…75
Humidity range without illumination [% RH] 10…80 10…80
Humidity fluctuation with illumination at 40 °C and 75 % RH [± % RH] 2 2
Humidity fluctuation with illumination at 25 °C and 60 % RH [± % RH] 1.5 1.5
Humidty recovery time after 30 s door open at 40 °C and 75 % RH with illumination [min] 6 6
Humidty recovery time after 30 s door open at 25 °C and 60 % RH with illumination [min] 4 4
Lightdata per Illumination Cassettes
ICH compliant illumination for photo stability testing [lx]* 7500 7500
ICH compliant illumination for photo stability testing [W/m²]* 1.1 1.1
Electrical data
Rated Voltage [V] 200…240 200…230
Power frequency [Hz] 50/60 50/60
Nominal power [kW] 2.4 2.4
Unit fuse [A] 16 16
Phase (Nominal voltage) 1~ 1~
  1. All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.
* Average value, measured at 25 °C with a spherical sensor (±10 %) by 12 cm below the light cassette. The values given in W/m2 refer to global radiation.

Accessories Series KBF LQC 240 | Humidity test chambers with ICH-compliant light source and light dose control