Designation |
KBFLQC720UL-240V¹ |
KBFLQC720-230V¹ |
Option model |
Standard |
Standard |
Order number |
9020-0335 |
9020-0334 |
Measures |
Interior volume [L] |
700 |
700 |
Net weight of the unit (empty) [kg] |
374 |
374 |
permitted load [kg] |
150 |
150 |
Load per rack [kg] |
45 |
45 |
Doors |
Inner doors |
2 |
2 |
Unit doors |
2 |
2 |
Internal Dimensions |
Width [mm] |
973 |
973 |
Height [mm] |
1250 |
1250 |
Depth [mm] |
576 |
576 |
Outer dimensions |
Width net [mm] |
1250 |
1250 |
Height net [mm] |
1925 |
1925 |
Depth net [mm] |
890 |
890 |
Wall clearance sidewise [mm] |
200 |
200 |
Wall clearance back [mm] |
100 |
100 |
Environment-specific data |
Energy consumption at 40 °C and 75 % RH [Wh/h] |
2350 |
2350 |
Sound-pressure level [dB(A)] |
53 |
53 |
Fixtures |
Number of shelves (std./max.) |
3/12 |
3/12 |
Number of illumination cassettes (std./max.) |
3/3 |
3/3 |
Performance Data Temperature |
Temperature range with 100% illumination [°C] |
10…60 |
10…60 |
Temperature range without illumination cassettes [°C] |
0…70 |
0…70 |
Max. heat compensation at 40 °C with illumination [W] |
1000 |
1000 |
Performance Data Climate |
Temperature range without illumination cassettes [°C] |
10…70 |
10…70 |
Temperature range with 100% illumination [°C] |
20…60 |
20…60 |
Temperature variation with illumination at 25 °C and 60 % RH [± K] |
1.2 |
1.2 |
Temperature variation with illumination at 40 °C and 75 % RH [± K] |
1.2 |
1.2 |
Temperature fluctuation with illumination at 40 °C and 75 % RH [± K] |
0.2 |
0.2 |
Temperature fluctuation with illumination at 25 °C and 60 % RH [± K] |
0.2 |
0.2 |
Humidity range without illumination [% RH] |
10…80 |
10…80 |
Humidity range with illumination cassettes, with illumination [% RH] |
10…75 |
10…75 |
Humidity fluctuation with illumination at 40 °C and 75 % RH [± % RH] |
2 |
2 |
Humidity fluctuation with illumination at 25 °C and 60 % RH [± % RH] |
2 |
2 |
Humidty recovery time after 30 s door open at 40 °C and 75 % RH with illumination [min] |
5 |
5 |
Humidty recovery time after 30 s door open at 25 °C and 60 % RH with illumination [min] |
1 |
1 |
Lightdata per Illumination Cassettes |
ICH compliant illumination for photo stability testing [lx]* |
9000 |
9000 |
ICH compliant illumination for photo stability testing [W/m²]* |
1.5 |
1.5 |
Electrical data |
Rated Voltage [V] |
200…240 |
200…230 |
Power frequency [Hz] |
50/60 |
50 |
Nominal power [kW] |
3.5 |
0 |
Unit fuse [A] |
16 |
16 |
Phase (Nominal voltage) |
1~ |
1~ |