• +2-35734687

  • sales@technoscientific.net

Stay Connected:

Series KT53 | Refrigerated incubators with thermoelectric refrigeration

code : KT53

Select version
Rated voltage Option model Version Art. No. MSRP*[?] Selection Bookmark
Model KT 53
200…240 V 1~ 50/60 Hz

Standard

KT053-230V¹ 9020-0311 - +
100…120 V 1~ 50/60 Hz

Standard

KT053UL-120V¹ 9020-0312 - +
* Recommended retail price

Select version
Rated voltage Option model Version Art. No. MSRP*[?] Selection Bookmark
Model KT 53
200…240 V 1~ 50/60 Hz

Standard

KT053-230V¹ 9020-0311 - +
100…120 V 1~ 50/60 Hz

Standard

KT053UL-120V¹ 9020-0312 - +
* Recommended retail price

 

Designation KT053UL-120V¹ KT053-230V¹
Option model Standard Standard
Order number 9020-0312 9020-0311
Measures
Interior volume [L] 53 53
Net weight of the unit (empty) [kg] 63 63
permitted load [kg] 40 40
Load per rack [kg] 15 15
Doors
Inner doors 1 1
Unit doors 1 1
Internal Dimensions
Width [mm] 400 400
Height [mm] 400 400
Depth [mm] 334 334
Outer dimensions
Width net [mm] 660 660
Height net [mm] 635 635
Depth net [mm] 630 630
Wall clearance back [mm] 100 100
Wall clearance sidewise [mm] 240 240
Environment-specific data
Energy consumption at 25 °C [Wh/h] 75 75
Energy consumption at 37 °C [Wh/h] 75 75
Sound-pressure level [dB(A)] 48 48
Fixtures
Number of shelves (std./max.) 2/5 2/5
Online
Shop URL https://www.binder-shop.com/us/incubators/refrigerated-incubators/refrigerated-incubator-with-thermoelectric-refrigeration-kt053ul-120v -
Performance Data Temperature
Temperature range [°C] 4…100 4…100
Temperature variation at 37 °C [± K] 0.3 0.3
Temperature fluctuation [± K] 0,1 ± K 0,1 ± K
Max. heat compensation at 37 °C [W] 100 100
Recovery time after 30 seconds door open at 37 °C [min] 2 2
Electrical data
Rated Voltage [V] 100…120 200…240
Power frequency [Hz] 50/60 50/60
Nominal power [kW] 0.4 0.4
Unit fuse [A] 10 10
Phase (Nominal voltage) 1~ 1~
  1. All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined in accordance to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.