• +2-35734687

  • sales@technoscientific.net

Stay Connected:

Series KT | Refrigerated incubators with thermoelectric refrigeration

code : KT170

Important characteristics

  • Temperature range: 4 °C to 100 °C
  • APT.line™ preheating chamber technology
  • Adjustable fan speed
  • Electric refrigeration with Peltier module
  • Controller with time-segment and real-time programming
  • Display via LCD monitor
  • Input via pushbutton/rotary knob
  • Inner door made of tempered safety glass
  • 2 stainless steel racks
  • Units up to 115 liters are stackable
  • Class 3.1 independent temperature safety device (DIN 12880) with visual and audible temperature alarm
  • Computer interface: Ethernet
  • Data recording and USB interface

Select version
Rated voltage Option model Version Art. No. MSRP*[?] Selection Bookmark
Model KT 170
200…240 V 1~ 50/60 Hz

Standard

KT170-230V¹ 9020-0289 - +
100…120 V 1~ 50/60 Hz

Standard

KT170UL-120V¹ 9020-0310 - +
* Recommended retail price

 

Designation KT170UL-120V¹ KT170-230V¹
Option model Standard Standard
Order number 9020-0310 9020-0289
Measures
Interior volume [cu.ft.] 5.8 5.8
Net weight of the unit (empty) [lbs] 225 225
permitted load [lbs] 265 265
Load per rack [lbs] 66 66
Doors
Inner doors 1 1
Unit doors 1 1
Internal Dimensions
Width [in] 23.64 23.64
Height [in] 30.14 30.14
Depth [in] 13.99 13.99
Outer dimensions
Width net [in] 33.88 33.88
Height net [in] 40.4 40.4
Depth net [in] 25.81 25.81
Wall clearance back [in] 3.94 3.94
Wall clearance sidewise [in] 9.46 9.46
Environment-specific data
Energy consumption at 37 °C [Wh/h] 80 80
Energy consumption at 25 °C [Wh/h] 80 80
Sound-pressure level [dB(A)] 48 48
Fixtures
Number of shelves (std./max.) 2/10 2/10
Performance Data Temperature
Temperature range [°C] 4…100 4…100
Temperature variation at 37 °C [± K] 0.4 0.4
Temperature fluctuation [± K] 0,1 ± K 0,1 ± K
Max. heat compensation at 40 °C [W] 100 100
Recovery time after 30 seconds door open at 37 °C [min] 3 3
Electrical data
Rated Voltage [V] 100…120 200…240
Power frequency [Hz] 50/60 50/60
Nominal power [kW] 0.8 0.8
Unit fuse [A] 10 10
Phase (Nominal voltage) 1~ 1~
  1. All technical data is specified for unloaded units with standard equipment at an ambient temperature of +22 °C ±3 °C and a power supply voltage fluctuation of ±10 %. The temperature data is determined to BINDER factory standard following DIN 12880, observing the recommended wall clearances of 10 % of the height, width, and depth of the inner chamber. Technical data refers to 100 % fan speed. All indications are average values, typical for units produced in series. We reserve the right to change technical specifications at any time.

Accessories Series KT | Refrigerated incubators with thermoelectric refrigeration